一种用于地面平面场近距测试的快速标定方法

    A Rapid Calibration Method for Near-field Testing in Ground Plane Range

    • 摘要: 在外场雷达散射截面积(RCS)测量中,通常使用同地相对标定法完成定标。但是,该方法需反复架设目标,操作过程繁琐,且频繁更换目标易导致测试背景改变,引入测量误差。针对这一问题,文中将远场RCS测试中使用的异地标定法引入到室外近距测试中。首先,分析了目标位于不同位置时的地平场增益与电磁波距离衰减;然后,通过实验测试,验证了地平场增益的影响因素;最后,利用现场采集的数据,通过矢量背景对消、快速傅里叶变换、频域加窗、地平场增益补偿与距离补偿等一系列方法,实现了对目标RCS的快速精确测量。实验结果表明,测量所得RCS均值误差在0.5 dB以内,RCS频域曲线与理论值的均方根误差均不超过1 dB。文中所提方法验证了异地标定法在室外近距RCS测试中具备可行性。

       

      Abstract: In field radar cross section (RCS) measurements, the on-site relative calibration method is typically used for calibration. However, this method requires repeatedly setting up the target, which is a cumbersome operation, and frequent target replacement can easily lead to changes in the test background, introducing measurement errors. To address this issue, the off-site calibration method used in far-field RCS testing has been introduced to outdoor close-range testing in this paper. First, ground-bounce illumination gain and the electromagnetic wave attenuation due to distance at different target positions are analyzed. Then, the factors influencing the ground-bounce illumination gain are verified by means of experimental testing. Finally, a rapid and accurate measurement of target′s RCS is achieved by utilizing on-site collected data and applying a series of methods, including background vector cancellation, Fourier transform, frequency-domain windowing, ground-bounce illumination gain compensation, and distance compensation. Experimental results show that the mean error of measured RCS falls within 0.5 dB, and the root mean square errrors of both the RCS curve in frequence domain and the theoretical vlaues fail to exceed 1 dB. The method proposed in the paper validates the feasibility of off-site calibration method in outdoor close-range RCS testing.

       

    /

    返回文章
    返回